VERIFICATION OF 235U, 238U BY MATHEMATICAL CALIBRATION CURVE FOR POINT SOURCES BY USING MONTE CARLO

Article

Last updated: 03 Jan 2025

Subjects

-

Tags

Physics

Abstract

The calibration of -ray spectrometer was performed by means of a set of certified standard point sources of energies that
cover the energy range of interest. The value of the absolute full energy peak efficiency "ab “ could be found by measuring
“CR" experimentally and the use of the known value of “ A “. Also, it is possible to calculate “ab" for a certain source to
detector distances (D) by using Monte Carlo [MC] simulation method.
The obtained results indicated that the fitting is identical to the experimental curve and MCNP calculations, that is both
these two methods (experimental and MCNP) can be used to extend the curve to cover the high energy range. It is possible to
drop any line energy inside the used energy range to estimate the absolute full energy peak efficiency, (for example, the energies
185.7, 1001 and 1275 keV corresponding to the sources 235U, 238U and 22Na). However, it can be estimated by using
any method from the previous methods (experimental or MCNP in addition to the fitting equation).

DOI

10.21608/absb.2014.22605

Keywords

point sources, experimental, fitting equation, MCNP, A mathematical calibration curve

Authors

First Name

S. A.

Last Name

Makhlouf

MiddleName

-

Affiliation

Faculty of Science, Al-Azhar University, Assiut, Egypt.

Email

-

City

-

Orcid

-

Volume

25

Article Issue

Issue 2-B

Related Issue

4289

Issue Date

2014-12-01

Receive Date

2014-10-17

Publish Date

2014-12-01

Page Start

9

Page End

14

Print ISSN

1110-2535

Online ISSN

2636-3305

Link

https://absb.journals.ekb.eg/article_22605.html

Detail API

https://absb.journals.ekb.eg/service?article_code=22605

Order

2

Type

Original Article

Type Code

520

Publication Type

Journal

Publication Title

Al-Azhar Bulletin of Science

Publication Link

https://absb.journals.ekb.eg/

MainTitle

VERIFICATION OF 235U, 238U BY MATHEMATICAL CALIBRATION CURVE FOR POINT SOURCES BY USING MONTE CARLO

Details

Type

Article

Created At

22 Jan 2023