Background: This paper aims to fill a research gap in the existing literature by investigating the impact of control, transient, and steady-state responses on the effectiveness of continuous positive airway pressure (CPAP) devices in the treatment of obstructive sleep apnea syndrome (OSAS).
Method: Bench tests were conducted on five CPAP devices: DreamStar Duo ST (SEFAM), Point 2 (Hoffrichter), Respironics PR System One Pro (Philips), iBreeze 20A (ResVent), and Respaps (MedTech). The evaluation focused on the transient responses of these devices to a pre-adjusted breathing pressure of 10 cmH2O. Parameters such as delay time, rise time, peak time, settling time, overshoot percentage, and steady-state error were measured and analyzed.
Results: The study revealed significant variations in the transient responses among the tested CPAP devices, attributed to the unique manufacturer algorithms. The observed differences underscore the importance of understanding and considering device-specific characteristics when selecting a CPAP device.
Conclusions: This research emphasizes the critical role of choosing the appropriate CPAP device for optimal patient outcomes, particularly for individuals experiencing breathing difficulties associated with OSAS. The findings highlight the need for careful consideration of device characteristics to ensure the effective treatment of sleep apnea using CPAP machines.

DOI

10.21608/jaet.2024.315504.1307

Keywords

CPAP device, sleep apnea, Step response characteristics, damped response

Authors

First Name

Mohamed

Last Name

Saad

MiddleName

Nagy

Affiliation

Biomedical Engineering Department, Faculty of Engineering, Minia University, Minia, Egypt

Email

m.n.saad@mu.edu.eg

City

Minia

Orcid

0000-0001-8229-0280

First Name

Nader

Last Name

Mohamed

MiddleName

A. Rahman

Affiliation

Biomedical Engineering Department, Faculty of Engineering, Misr University for Science and Technology (MUST), Giza 12568, Egypt

Email

nader.shaaban@must.edu.eg

City

Cairo

Orcid

0000-0001-7680-306X

First Name

Shrief

Last Name

Abdelazeez

MiddleName

-

Affiliation

Systems and Biomedical Engineering Department, Faculty of Engineering, Cairo University, Giza, Egypt. Nalecz Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Science, Poland

Email

shrief.s.abdelazeez@eng1.cu.edu.eg

City

Cairo

Orcid

-

First Name

Abd Elrhman

Last Name

Shaaban

MiddleName

-

Affiliation

Biomedical Engineering Department, Higher Institute of Engineering, El-Shorouk Academy, Cairo, Egypt

Email

a.shaaban@sha.edu.eg

City

Cairo

Orcid

-

First Name

Wael

Last Name

Abouelwafa

MiddleName

-

Affiliation

Biomedical Engineering Department, Faculty of Engineering, Minia University, Minia, Egypt

Email

wael.wafa@mu.edu.eg

City

Cairo

Orcid

-

Volume

44

Article Issue

1

Related Issue

53703

Issue Date

2025-01-01

Receive Date

2024-08-26

Publish Date

2025-01-01

Page Start

194

Page End

200

Print ISSN

2682-2091

Online ISSN

2812-5487

Link

https://jaet.journals.ekb.eg/article_414261.html

Detail API

http://journals.ekb.eg?_action=service&article_code=414261

Order

414,261

Type

Original Article

Type Code

1,142

Publication Type

Journal

Publication Title

Journal of Advanced Engineering Trends

Publication Link

https://jaet.journals.ekb.eg/

MainTitle

Unmasking Effectiveness: A Comprehensive Analysis of CPAP Devices in Treating Sleep Apnea

Details

Type

Article

Created At

09 Mar 2025