Abstract Background: Quality Assurance (QA) in radiotherapy is an important studying issue in order to test the performance of medical Linac machine and verify the delivering accuracy of patient treatment plan. Aim of Study: Clinical validation of using compass system in VMAT QA for abdomen site pediatric patients. Material and Methods: Ten VMAT neuroblastoma plans were created using Monaco TPS. CT images, RT Structures, RT Plans, and RT doses for all cases were exported to COMPASS system through DICOM network then COMPASS computations were performed using collapsed cone algorism for all patients with a calculation grid resolution of 2mm from the imported DICOM RT plans. Finally, a comparison was performed based on dose difference and all VMAT plans were measured at two matched linear accelerators. Results: The mean dose of PTV and D99% for the ten cases were measured. The comparison between the Monaco dose calculation, compass computed dose and compass reconstructed dose showed a good agreement within maximum deviation 4% for the PTVs and critical organs in point doses and volume doses. Conclusions: Compass system is a reliable effective system to perform 3D dose verification of VMAT plans for neuroblastoma patients. So Compass QA system can be used as an effective tool before VMAT treatment. The variations of dose between TPS and Compass were due to mode of calculation (Compass algorism uses dose to water for calcu-lation while Monte-Carlo algorism uses dose to medium).

DOI

10.21608/mjcu.2022.296223

Keywords

QA, VMAT, TPS, COMPASS, PTV

Authors

First Name

SAMAR E. MOHAMED, M.Sc.*;

Last Name

NOHA G. MADIAN, Ph.D.**

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

REEM H. EL-GEBALY, Ph.D.**;

Last Name

HANI AMMAR, Ph.D.***,*

MiddleName

-

Affiliation

The Department of Biophysics, Children Cancer Hospital 57357*, Biophysics Department, Faculty of Science, Cairo University** and Clinical Oncology Department, Faculty of Medicine, Aswan University***

Email

-

City

-

Orcid

-

Volume

90

Article Issue

12

Related Issue

39806

Issue Date

2022-12-01

Receive Date

2023-04-23

Publish Date

2022-12-01

Page Start

2,731

Page End

2,737

Print ISSN

0045-3803

Online ISSN

2536-9806

Link

https://mjcu.journals.ekb.eg/article_296223.html

Detail API

https://mjcu.journals.ekb.eg/service?article_code=296223

Order

296,223

Type

Original Article

Type Code

263

Publication Type

Journal

Publication Title

The Medical Journal of Cairo University

Publication Link

https://mjcu.journals.ekb.eg/

MainTitle

Clinical Validation of the COMPASS System for Two Matched Elekta Versa HD Linear Accelerators in Neuroblastoma

Details

Type

Article

Created At

30 Dec 2024