Abstract

Purpose
The purpose of this study is to evaluate the effect of customized acquisition parameters on image quality in cardiac perfusion imaging using the SPECT/CT Symbia Intevo Excel system. Specifically, it aims to assess how changes in matrix size, number of views, and acquisition time impact the Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR) in cardiac images.

Materials and Methods
This study used the SPECT/CT Symbia Intevo Excel imaging system with a 128x128 matrix, acquiring 32 views with 30 seconds per view. To enhance image quality, the Butterworth filter was applied. The SNR and CNR values were calculated to quantitatively assess image quality with a series of test images obtained under these customized acquisition settings.

Results
The customized acquisition parameters resulted in a significant improvement in both SNR and CNR compared to standard settings. These optimized parameters led to clearer and more detailed images, enhancing the visibility of cardiac structures and abnormalities. Statistical analysis validated these improvements, showing a significant enhancement in image quality.

Conclusion
This study demonstrates that customizing acquisition parameters in the Symbia Intevo Excel SPECT/CT system can significantly improve image quality in cardiac perfusion imaging. The findings highlight the importance of optimizing acquisition protocols to improve diagnostic outcomes and patient care. Further research is suggested to explore additional factors impacting image quality and to help establish standardized protocols in cardiac imaging.

DOI

10.21608/aujst.2024.327603.1131

Keywords

qualitative analysis, phantom cardiac, Image processing, Nuclear medicine

Authors

First Name

Eman

Last Name

ebraheem

MiddleName

Amreen

Affiliation

physics, faculty of science, aswan university

Email

emanamreenebraheem@gmail.com

City

aswan

Orcid

-

First Name

Kassem

Last Name

Behairy

MiddleName

O

Affiliation

Department of Physics, Faculty of Sciences, Aswan University

Email

kass_phys@aswu.edu.eg

City

aswan

Orcid

-

First Name

Jamal

Last Name

Yahia

MiddleName

-

Affiliation

Department of Physics, Aswan Faculty of Sciences, Aswan University

Email

gamal102@aswu.edu.eg

City

-

Orcid

-

Volume

4

Article Issue

4

Related Issue

51586

Issue Date

2024-12-01

Receive Date

2024-10-11

Publish Date

2024-12-01

Page Start

119

Page End

133

Print ISSN

2735-3087

Online ISSN

2735-3095

Link

https://aujst.journals.ekb.eg/article_391911.html

Detail API

https://aujst.journals.ekb.eg/service?article_code=391911

Order

391,911

Type

Original papers

Type Code

2,312

Publication Type

Journal

Publication Title

Aswan University Journal of Sciences and Technology

Publication Link

https://aujst.journals.ekb.eg/

MainTitle

Cardiac SPECT Image Quality Comparison: Customized vs. Siemens Using Phantom

Details

Type

Article

Created At

29 Dec 2024