Abstract

Purpose
To compare and assess the accuracy of both Lenstar T-cone toric platform and Verion in toric intraocular lens (IOL) power calculation regarding postoperative refraction.
Design
The study design was a retrospective one.
Participants and methods
A total of 33 eyes (33 patients) with astigmatism at least 1 D underwent cataract surgery. The eyes included in the study were divided into two groups. Group 1 included 17 eyes where IOL was implanted according to Verion IOL power calculations, and group 2 included 16 eyes where IOL was implanted according to Lenstar T-cone power calculations. The keratometric readings of the 33 eyes of both devices were compared. Postoperative manifest refraction data of all patients that were taken 1 month postoperative were collected. The manifest postoperative refraction values of both groups were compared with both the expected refraction predicted by Lenstar toric T-cone and with that of the Verion.
Results
There were no statistical differences between the two devices used in this study regarding keratometric values. There were no statistical differences between the two devices used in this study regarding neither the final residual postoperative spherical refractive error nor the final residual postoperative cylindrical refractive error, nor the final residual postoperative spherical equivalent.
Conclusion
Both the Lenstar LS 900 T-cone biometer and the Verion image-guided system provide excellent, accurate, reproducible, and comparable postoperative results.

DOI

10.4103/JCRS.JCRS_5_19

Keywords

LENSTAR, toric intraocular lens power calculation, VERION

Authors

First Name

Ahmed M.

Last Name

Eid

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

Volume

24

Article Issue

2

Related Issue

49505

Issue Date

2018-07-01

Receive Date

2019-03-12

Publish Date

2018-07-01

Page Start

47

Page End

52

Print ISSN

1687-6997

Online ISSN

2536-9725

Link

https://ejcrs.journals.ekb.eg/article_370250.html

Detail API

https://ejcrs.journals.ekb.eg/service?article_code=370250

Order

370,250

Publication Type

Journal

Publication Title

The Egyptian Journal of Cataract and Refractive Surgery

Publication Link

https://ejcrs.journals.ekb.eg/

MainTitle

Comparison between LENSTAR and VERION in toric intraocular lens power calculation

Details

Type

Article

Created At

28 Dec 2024