Purpose
The aim of this study was to assess white-to-white (WTW) horizontal corneal diameter using Pentacam versus IOLMaster and determine whether the two instruments were interchangeable.
Patients and methods
Twenty eyes were included in this prospective clinical study conducted in I-care Eye Centre, Alexandria. Measurement of WTW horizontal corneal diameter was taken using Pentacam (internal and external measurements) and IOLMaster.
Results
The mean WTW measured using IOLMaster was 11.66 ± 0.27 mm, ranging from 11.30 to 12.30 mm. The mean WTW externally using Pentacam was 11.93 ± 0.43 mm, ranging from 11.01 to 12.66 mm. The mean WTW internally using Pentacam was 11.20 ± 0.39 mm, ranging from 10.56 to 11.93 mm. The diameters measured using IOLMaster were significantly less than that measured using Pentacam externally and significantly more than that measured using Pentacam internally. There was no correlation between WTW measured using the two machines.
Conclusion
Although horizontal corneal diameter (WTW) can be measured using Pentacam or IOLMaster, WTW measurement is significantly different between the two instruments and they should not be used interchangeably. Pentacam measurement depends on manual placement of calipers on the Scheimpflug digital image.

DOI

10.4103/1687-6997.194365

Keywords

IOLMaster, pentacam, white-to-white

Authors

First Name

Mohammed W.

Last Name

Elkateb

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

Hesham S.

Last Name

Swelem

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

Volume

22

Article Issue

1

Related Issue

49498

Issue Date

2016-01-01

Receive Date

2016-03-10

Publish Date

2016-01-01

Page Start

10

Page End

14

Print ISSN

1687-6997

Online ISSN

2536-9725

Link

https://ejcrs.journals.ekb.eg/article_370223.html

Detail API

https://ejcrs.journals.ekb.eg/service?article_code=370223

Order

370,223

Publication Type

Journal

Publication Title

The Egyptian Journal of Cataract and Refractive Surgery

Publication Link

https://ejcrs.journals.ekb.eg/

MainTitle

Measurement of White-to-White Distance Using Pentacam Scheimpflug Imaging versus IOLMaster

Details

Type

Article

Created At

28 Dec 2024