Beta
357614

Evaluation of surface roughness and fracture resistance of maxillary conventional & digital complete dentures: An Invitro study

Article

Last updated: 04 Jan 2025

Subjects

-

Tags

Prosthodontics

Abstract

Aim: to evaluate the surface roughness and fracture resistance of complete dentures fabricated using heat cured polymethyl methacrylate (PMMA) resin manufactured by conventional technique & digital complete dentures manufactured by additive manufacturing technique. Materials and Methods: A total of 24 complete dentures were fabricated (n=24), 12 dentures were fabricated using heat cured polymethyl methacrylate (PMMA) resin manufactured by the conventional methods. 12 dentures were designed by Exocad software program & constructed using additive manufacturing technique. Surface roughness was evaluated using optical profilometry. Fracture resistance was evaluated using universal testing machine. Data were analysed using unpaired t-test where α = 0.05. Results: Concerning surface roughness, there was no statistically significant difference between the two groups (P=0.5137). Conventional PMMA dentures showed mean (2.52± 0.14 µm), while 3D-printed dentures showed mean (2.59± 0.38 µm). However, regarding the fracture resistance, there was statistically significant difference between the two groups (P<0.5), the conventional PMMA dentures showed higher mean (572.49± 13.07 N), while the digital dentures showed lesser mean with (202.51 ± 10.35 N). Conclusion: Conventional PMMA dentures and 3D-printed dentures had similar surface roughness. However conventional PMMA dentures had better fracture resistance when compared to 3D-printed dentures. KEYWORDS: Digital dentures, fracture resistance, surface roughness, complete dentures.

DOI

10.21608/asdj.2024.277269.1231

Keywords

Digital dentures, Fracture resistance, Surface roughness, Complete dentures, Exocad software

Authors

First Name

Eman

Last Name

Abdelghaffar

MiddleName

Gamal

Affiliation

Removable Prosthodontics Department, Faculty of Dentistry, The British University in Egypt

Email

eman.khedr@bue.edu.eg

City

-

Orcid

0009-0007-6956-8293

Volume

34

Article Issue

2

Related Issue

48077

Issue Date

2024-06-01

Receive Date

2024-03-17

Publish Date

2024-06-01

Page Start

42

Page End

48

Print ISSN

1110-7642

Online ISSN

2735-5039

Link

https://asdj.journals.ekb.eg/article_357614.html

Detail API

https://asdj.journals.ekb.eg/service?article_code=357614

Order

6

Type

Original articles

Type Code

1,638

Publication Type

Journal

Publication Title

Ain Shams Dental Journal

Publication Link

https://asdj.journals.ekb.eg/

MainTitle

Evaluation of surface roughness and fracture resistance of maxillary conventional & digital complete dentures: An Invitro study

Details

Type

Article

Created At

28 Dec 2024