Introduction: Cochlear implant (CI) has been widely used in cases with severe to profound sensorineural hearing loss. The integrity of the internal CI part after implantation could be objectively assessed by measuring the auditory response telemetry (ART) based on each individual electrode's ECAP threshold, and by postoperative imaging which provides detailed information on the position of individual electrodes specially with measurement of the electrode- modiolus distance (EMD).
Objective: We aimed to correlate between the auditory response telemetry (ART) measurements and radiological position of the electrode arrays (Electrode-Modiolus distance) inside the cochlea.
Materials and methods: Twenty-five children with severe to profound sensorineural hearing loss who had implanted unilaterally with MED-EL SONATA FLEX 28 were enrolled. Intraoperative ART results for each array were obtained. Postoperative imaging by means of a128-Multidetector CT scanner with sagittal and oblique multiplanar reconstructions (MPRs) were obtained. Images were analyzed to measure the electrode modiolus distance (EMD) and results were correlated with ART measurements.
Results: The mean ART threshold for each electrode ranged between 15.97±4.69 qu and 20.61±6.64 qu. While, the mean electrode-modiolus distance for each electrode fluctuated from 1.72±0.38 mm to 4.53± 1.27 mm. Pearson test showed a statistically significant positive correlation between the electrode- modiolus distance and the auditory response telemetry (ART) measurements.
Conclusion: Postoperative computerized tomography scanning could provide a valuable tool to assess the outcome of cochlear implant through measuring the electrode modiolus distance (EMD).
DOI
10.21608/ejnso.2023.292165
Keywords
Auditory response telemetry, Electrode modiolus distance, Cochlear implant
Authors
Affiliation
Otorhinolaryngology department Assiut university, Assiut University, Assiut, Egypt.
Email
samahhamdy752@gmail.com
City
-Orcid
-Affiliation
Otorhinolaryngology Department, Assiut university, Assiut University, Assiut, Egypt
Email
ahmedaleem56@yahoo.com
Orcid
-Affiliation
Radiology Department. , Faculty of medicine Assiut university , Assuit ,Egypt
Email
hazem_yousef71@aun.edu.eg
Orcid
-MiddleName
-Affiliation
Otolaryngology Department, Assuit University , Assiut, Egypt
Email
m.aboshanif@aun.edu.eg
Orcid
-Link
https://ejnso.journals.ekb.eg/article_292165.html
Detail API
https://ejnso.journals.ekb.eg/service?article_code=292165
Publication Title
Egyptian Journal of Neck Surgery and Otorhinolaryngology
Publication Link
https://ejnso.journals.ekb.eg/
MainTitle
Correlation between auditory response telemetry (ART) and computerized tomography measurements in detection of cochlear implant electrode position