Beta
367461

Retention force of two different pickup strategies for Locator retained mandibular implant overdenture; an in-vitro study.

Article

Last updated: 24 Dec 2024

Subjects

-

Tags

Prosthodontics

Abstract

This article evaluates the retention force of Locator metal housing attached to implant overdenture with two different pickup strategies.
Two dummy implants were inserted in the lower canine positions. The denture base was designed digitally with particular parameters for both study groups (n = 5 each) with two different pickup strategies: without retaining material for the intervention group and with retaining material for the control group. Retention forces were tested by using the universal testing machine at 0, 135, 270, 405 and after 540 insertion/separation cycles respectively.
The data was analyzed to evaluate the effect of time on retention in each study group using the One-Way ANOVA test, which revealed an insignificant difference between different intervals. To compare the two study groups with each other Independent t-test was used which revealed insignificant differences between them in all intervals. Frequency and percentages analysis of detached and non-detached metal housings in both groups was performed by using Chi-square test which also revealed insignificant differences between both groups.
Within the limitations of this study, the retention in both groups was nearly equivalent. Yet, by time, retention decreased in both study groups. This pickup technique may be a new application of digital dentistry reducing chair-side corrections.

DOI

10.21608/adjc.2023.239566.1400

Keywords

Implant, edentulism, Locator attachment, Over denture pickup, digital dentistry

Authors

First Name

Noha

Last Name

Barakat

MiddleName

Waleed

Affiliation

Department of Prosthodontics, Faculty of Dentistry, Cairo University, Giza, Egypt.

Email

noha.barakat@dentistry.cu.edu.eg

City

-

Orcid

0009-0002-0600-9412

First Name

Mahmoud

Last Name

El Far

MiddleName

Mokhtar

Affiliation

Department of Prosthodontics, Faculty of Dentistry, Cairo University, Giza, Egypt.

Email

mahmoud.elfar@dentistry.cu.edu.eg

City

Cairo

Orcid

-

First Name

Ahmed

Last Name

Kotb

MiddleName

Mustafa

Affiliation

Department of Prosthodontics, Faculty of Dentistry, Cairo University, Giza, Egypt.

Email

ahmed.kotb@dentistry.cu.edu.eg

City

Giza

Orcid

-

Volume

6

Article Issue

3

Related Issue

48861

Issue Date

2024-07-01

Receive Date

2023-09-29

Publish Date

2024-07-01

Page Start

547

Page End

558

Print ISSN

2636-302X

Online ISSN

2636-3038

Link

https://adjc.journals.ekb.eg/article_367461.html

Detail API

https://adjc.journals.ekb.eg/service?article_code=367461

Order

367,461

Type

Original Article

Type Code

688

Publication Type

Journal

Publication Title

Advanced Dental Journal

Publication Link

https://adjc.journals.ekb.eg/

MainTitle

Retention force of two different pickup strategies for Locator retained mandibular implant overdenture; an in-vitro study.

Details

Type

Article

Created At

24 Dec 2024