Aim: The aim of this study was to compare the accuracy of implant placement when using tooth supported surgical guide versus tooth tissue supported one manufactured by DLP (Digital Light Processing) technique.
Methodology: 21replica implants were inserted in 7 partially edentulous maxillary resin cast. Surgical implant placement was done using static DLP printed surgical guides, Three designs of surgical guides were designed representing different type and position of support (Anterior tooth-supported, Posterior tooth-supported and posterior tooth-tissue supported) respectively when using the same number of tooth support.
. After implant placement Accuracy measurements were done using (blue sky bio software) by superimposing the actually placed implants with the virtually planned implants.
Results : Regarding mesiodistal and buccolingual distance, lack of statistical significant difference was found between both group 1 and 2 while both show stistical significant difference in compare with (tooth tissue supported ones ) .
Regarding the depth results. Post tooth supported guides show a statistical significant difference over both group 1 and group 3 While the ant tooth supported guide was in statistical significant with tooth tissue supported one.
Conclusion: using 3 units of posterior tooth supported guides offer accurate implant positioning. on the other hand, using Three units anterior tooth supported surgical guides have shown less yet still satisfactory results of accuracy when compared to posterior tooth supported ones , using3 units of tooth tissue supported guides results in higher significant deviation value thus it was risky to be used even In single implant placement.

DOI

10.21608/adjc.2023.182789.1220

Keywords

surgical guides, Implants, Support, accuracy

Authors

First Name

John

Last Name

Isaac

MiddleName

Zakaria

Affiliation

Prosthodontic department cairo university

Email

john.zakaria@dentistry.cu.edu.eg

City

-

Orcid

0000-0003-4254-966X

First Name

Marwa

Last Name

El Sadek

MiddleName

Abd elAal

Affiliation

Prosthodontic department, cairo university

Email

marwa.abdelaal1@gmail.com

City

cairo

Orcid

0000-0003-3541-3254

First Name

Mohamed

Last Name

Nada

MiddleName

Abdelmonem

Affiliation

Prosthodontic department , cairo university

Email

mohamed.nada@dentistry.cu.edu.eg

City

cairo

Orcid

-

Volume

5

Article Issue

3

Related Issue

42407

Issue Date

2023-07-01

Receive Date

2022-12-23

Publish Date

2023-07-01

Page Start

627

Page End

638

Print ISSN

2636-302X

Online ISSN

2636-3038

Link

https://adjc.journals.ekb.eg/article_316509.html

Detail API

https://adjc.journals.ekb.eg/service?article_code=316509

Order

316,509

Type

Original Article

Type Code

688

Publication Type

Journal

Publication Title

Advanced Dental Journal

Publication Link

https://adjc.journals.ekb.eg/

MainTitle

Accuracy of Implant Placement With Tooth Versus Tooth Tissue Supported Digital Light Processing Guides: Non Randomized In-vitro Study

Details

Type

Article

Created At

24 Dec 2024