Aim: The aim of this in vitro study was to evaluate the accuracy of scanning stock abutment versus scan bodies on a single implant using two different intra-oral scanners.
Methodology: A printed cast model of a partially edentulous maxilla installed with a single implant at the second premolar area was fabricated to serve as reference cast. Two scannable components were used in this study; a scan body (control group) and a stock abutment (intervention group). For both groups, the cast was scanned once with an extra-oral scanner and five times with two different intraoral scanners; Medit i700 and Prime scan. Geomagic X- software was used to compare the trueness and precision in the two groups. Precision was recorded twice (with full model scans and after segmentation of the scannable components from their corresponding models). RMS values were recorded, tabulated and analyzed.
Results: There were no statistically significant differences in trueness between the 2 groups (scan body and stock abutment groups) for both intraoral scanners. There was no statistically significant difference in trueness between the two intraoral scanners within each group. Primescan IOS showed significantly lower precision than Medit i700 in the full model scans of the scan body group.
Conclusions: Trueness was significantly comparable for both groups and was not significantly affected by the type of IOS used. However, precision of Medit i700 was higher than that of Primescan IOS when scan bodies were used especially in the full model scans.

DOI

10.21608/edj.2024.312589.3166

Keywords

Trueness, stock abutments, scan bodies, Intraoral Scanners, Precision

Authors

First Name

Ahmad

Last Name

Abodheir

MiddleName

Saad

Affiliation

MSc Student, Master candidate, Masters of implantology, Faculty of Dentistry, Cairo University

Email

ahmad.abodheir@dentistry.cu.edu.eg

City

-

Orcid

0009-0004-9478-2185

First Name

Nermeen

Last Name

Hassan

MiddleName

Ahmed

Affiliation

Lecturer of Prosthodontics, Faculty of Dentistry, Cairo University

Email

nermeen.hassan@denistry.cu.edu.eg

City

-

Orcid

0000-0002-5393-9607

First Name

Mohamed

Last Name

Abdallah

MiddleName

Farouk

Affiliation

Professor, Department of Prosthodontics, Faculty of Dentistry, Cairo University; Professor, Department of Prosthodontics, Faculty of Dentistry, Galala University, Egypt

Email

mohamed.abdalla@dentistry.cu.edu.eg

City

-

Orcid

0000-0002-6617-9379

First Name

Amal

Last Name

Swelem

MiddleName

Ali

Affiliation

Professor of Prosthodontics, Faculty of Dentistry, Cairo University

Email

amal.swelem@dentistry.cu.edu.eg

City

cairo

Orcid

0000-0001-5658-1586

Volume

70

Article Issue

4

Related Issue

50687

Issue Date

2024-10-01

Receive Date

2024-08-17

Publish Date

2024-10-01

Page Start

3,711

Page End

3,720

Print ISSN

0070-9484

Online ISSN

2090-2360

Link

https://edj.journals.ekb.eg/article_383249.html

Detail API

https://edj.journals.ekb.eg/service?article_code=383249

Order

383,249

Type

Original Article

Type Code

254

Publication Type

Journal

Publication Title

Egyptian Dental Journal

Publication Link

https://edj.journals.ekb.eg/

MainTitle

Accuracy of scanning stock abutments versus scan bodies using different intraoral scanners for single implants. An in vitro study

Details

Type

Article

Created At

23 Dec 2024