Abstract

The aim of the study is to compare stresses induced onto implant-bone interface, implants & denture bases on using cobalt Chromium (Cr-Co) and Titanium (Ti) framework materials.
Methodology:
Three dimensional finite element models of completely edentulous maxilla rehabilitated with ALL on 4 implant prostheses, (Two vertical anterior implants & two posterior implants 20°,30°, 45° mesially tilted ) , Cr-Co & Ti framework materials were constructed. Bilateral axial load of 100 N was applied onto the palatal cusps of maxillary posterior teeth. The resultant equivalent Von Misses stresses at areas of interest were calculated.
Results:
The Von Mises stresses induced in bone were 5.132 & 6.144 Mpa in model I & IV / 2.403 & 2.272 Mpa in model II & V /13.937 & 14.591 Mpa in model III & VI respectively .
The stresses induced in denture bases were 225.25 & 218.91 Mpa in model I & IV, 156.74 & 170.19 Mpa in model II & V, 227.25 & 159.99 Mpa in model III & VI respectively.
The stresses induced in implants were 26.591 & 31.315 Mpa in model I & IV, 18.322 & 26.42 Mpa in model II & V / 117.29 & 117.66 Mpa in model III & VI respectively.
Conclusions:
Within the limitations of this study, we may conclude:
• The more rigid the framework, the less stress dissipation onto the surrounding structures.
• Posterior implants tilted mesially may reduce the stresses induced onto implants & surrounding structures.

DOI

10.21608/edj.2023.197760.2470

Keywords

All on four, maxilla, FEA, implant tilting, Framework

Authors

First Name

Mohamed

Last Name

Hakim

MiddleName

-

Affiliation

Associate Professor of Removable Prosthodontics, Faculty of Dental Medicine, Beni-Suef University

Email

mabdelhakim@yahoo.com

City

Giza

Orcid

0000-0001-9349-7885

Volume

69

Article Issue

2

Related Issue

40704

Issue Date

2023-04-01

Receive Date

2023-03-03

Publish Date

2023-04-01

Page Start

1,517

Page End

1,527

Print ISSN

0070-9484

Online ISSN

2090-2360

Link

https://edj.journals.ekb.eg/article_294292.html

Detail API

https://edj.journals.ekb.eg/service?article_code=294292

Order

32

Type

Original Article

Type Code

254

Publication Type

Journal

Publication Title

Egyptian Dental Journal

Publication Link

https://edj.journals.ekb.eg/

MainTitle

Three Dimensional Finite Element Analysis of Stresses Induced Around Inclined Implants with Different Framework Materials in Maxillary Complete Arch Prostheses (All on 4)

Details

Type

Article

Created At

23 Dec 2024