Background
I-gel is designed to suit the anatomy of hypopharyngeal and perilaryngeal areas in adults without an inflatable cuff. There is insufficient evidence regarding quality of seal of I-gel during PEEP application in pediatric patients. The objective of this study was to evaluate the performance of I-gel usage in children during general anesthesia with PEEP application at a level of 5 cm HO and assess whether it improves oxygenation.
Patients and methods
A total of 42 ASA physical status I, and II children undergoing surgery under general anesthesia were included. Patients were randomly allocated to one of two equal groups to be on Pressure-control ventilation (PCV) with PEEP 5 cm HO (group I) and PCV without PEEP (group II). I-gel size 2 1/2 was used in children weighing from 25 to 35 kg. Leak volume (LV) and leak fraction (LF) were recorded. Peak Inspiratory Pressure (PIP), expiratory and inspiratory tidal volume as well as minute volume and End tidal CO (ETCO) were also recorded at 5 min, 30 min, and 1 h after I-gel insertion.
Results
Leak volume and leak fraction had no statistical significant differences between both groups. Patients with PEEP had significantly lower (ETCO), higher PIP, higher inspiratory tidal volume, and higher expiratory tidal volume (p = 0.001) during the post I-gel insertion follow up period. Patients with PEEP also had significantly higher PaO and lower PaCO levels (p = 0.001).
Conclusions
I-gel may be used safely during PCV while applying PEEP of 5 cm HO in children with an effective seal pressure, improvement in oxygenation and without leak or gastric insufflation.

DOI

10.1016/j.egja.2018.04.003

Authors

First Name

Ghada

Last Name

Kamhawy

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

Ahmed

Last Name

El-Lilly

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

Mona

Last Name

Azzam

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

Maha

Last Name

Misk

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

Volume

34

Article Issue

3

Related Issue

51206

Issue Date

2018-07-01

Receive Date

2017-11-24

Publish Date

2018-07-01

Page Start

101

Page End

105

Print ISSN

1110-1849

Online ISSN

1687-1804

Link

https://egja.journals.ekb.eg/article_388564.html

Detail API

https://egja.journals.ekb.eg/service?article_code=388564

Order

388,564

Publication Type

Journal

Publication Title

Egyptian Journal of Anaesthesia

Publication Link

https://egja.journals.ekb.eg/

MainTitle

Positive end-expiratory pressure with I-gel in children, is it effective and safe?

Details

Type

Article

Created At

21 Dec 2024