Abstract

Background
Cardiac output is an important determinant of tissue perfusion, with several methods described to assess it. Electrical cardiometry is a new noninvasive method for determination of stroke volume variation (SVV), which is used to calculate the cardiac output.
Aim of work
This study aimed to validate the electrical cardiometry measurements of SVV compared with measurements of SVV taken by transesophageal echocardiography (TEE).
Methods
A total of 38 patients were included in this study. Hemodynamic parameters were obtained by TEE and cardiometry soon after closure of the sternum and after volume expansion.
Results
SVV after 10 min of sternum closure by TEE was 15.5% (SD=7.1), and SVV after 10 min of sternum closure by cardiometry was 14.3% (SD=6.1). Bland–Altman analysis revealed a mean bias of −1.2. The 1.96 SD limits of agreement were −8 to 5.7%.
Conclusion
There is a good correlation between SVV measured by TEE and that measured by cardiometry. Cardiometry can be used as a noninvasive hemodynamic monitoring in patients undergoing coronary artery bypass surgery surgery.

DOI

10.4103/ejca.ejca_8_22

Keywords

CARDIOMETRY, Stroke Volume Variation, transeosphageal echocardiography

Authors

First Name

Ahmed

Last Name

Said

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

Maged

Last Name

Salah

MiddleName

-

Affiliation

-

Email

maged.salah@kasralainy.edu.eg

City

-

Orcid

-

First Name

Sherif

Last Name

Mamdouh

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

Eslam

Last Name

Heggy

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

Mohamed

Last Name

Wagih

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

Volume

16

Article Issue

3

Related Issue

48393

Issue Date

2023-01-01

Receive Date

2022-07-03

Publish Date

2023-01-03

Print ISSN

1687-9090

Online ISSN

2090-326X

Link

https://ejca.journals.ekb.eg/article_359715.html

Detail API

https://ejca.journals.ekb.eg/service?article_code=359715

Order

359,715

Publication Type

Journal

Publication Title

The Egyptian Journal of Cardiothoracic Anesthesia

Publication Link

https://ejca.journals.ekb.eg/

MainTitle

Validation of stroke volume variation assessed by electrical cardiometry to predict fluid responsiveness in patients undergoing coronary artery bypass surgery after closure of the sternum: an observational study

Details

Type

Article

Created At

20 Dec 2024