Abstract

Objective: to compare the acoustic output of conventional two-dimensional US and 3D/4D US in a 
group of pregnant women with fetal craniofacial malformations. Method: Fifty women with 
suspected craniocerebral anomalies during routine prenatal ultrasound examination are included in the 
study. For ultrasound assessment, we used a Doppler ultrasound system. Subsequently, 3D/4D images 
were obtained using a suitable probe. During both examinations, changes in thermal index (TI) and 
mechanical index (MI) were recorded. Results: Comparison between TI and MI of the different 
ultrasound techniques revealed no statistically significant differences. The reported TI for the studied 
modalities was 0.277 ± 0.01, 0.271 ± 0.008 and 0.249 ± 0.01 (p=0.43) while MI was 1.12 ± 0.012, 
0.87 ± 0.014 and 1.19 ± 0.37 (p=11) for B mode, 3D and 4D techniques respectively. Conclusion:
3D/4D ultrasound imaging is comparable to 2D imaging regarding thermal and mechanical indexes in 
women with fetal craniocerebral anomalies. 

DOI

10.21608/mjmr.2022.221604

Keywords

Craniocerebral anomalies, 3D/4D ultrasound, thermal index, mechanical index

Authors

First Name

Mustafa.

Last Name

Osman

MiddleName

A. A

Affiliation

Department of Obstetrics and Gynecology, El-Hussein University Hospital, Al-Azhar University

Email

-

City

-

Orcid

-

First Name

Mohamed

Last Name

Shehata

MiddleName

S. A.

Affiliation

Department of Radiology, El-Hussein University Hospital, Al-Azhar University, Cairo, Egypt

Email

-

City

-

Orcid

-

First Name

Khaled

Last Name

Abd-Elaziz

MiddleName

M. S.

Affiliation

Department of Pediatrics, El-Hussein University Hospital, Al-Azhar University, Cairo, Egypt

Email

-

City

-

Orcid

-

First Name

Mohamed

Last Name

Rashad

MiddleName

H.

Affiliation

Department of Neurology, El-Hussein University Hospital, Al-Azhar University, Cairo, Egypt

Email

-

City

-

Orcid

-

Volume

30

Article Issue

4

Related Issue

31649

Issue Date

2019-10-01

Receive Date

2022-02-24

Publish Date

2019-10-01

Page Start

20

Page End

22

Online ISSN

2682-4558

Link

https://mjmr.journals.ekb.eg/article_221604.html

Detail API

https://mjmr.journals.ekb.eg/service?article_code=221604

Order

5

Type

Original Article

Type Code

2,212

Publication Type

Journal

Publication Title

Minia Journal of Medical Research

Publication Link

https://mjmr.journals.ekb.eg/

MainTitle

Output Indices in 2D versus 3D/4D Ultrasonography in Fetal Craniocerebral Anomalies

Details

Type

Article

Created At

23 Jan 2023