In this work, a nuclear gauge, based on single and dual energy γ-ray
transmissions, has been developed and used to determine the concentrations of
the aluminum component in copper-aluminum alloy. Disk foils of 7.07 cm2 area
and thicknesses ranging from 0.48 to 0.834 cm have been prepared from the
alloy to be used as targets for the γ-ray photons. Criteria for the best choice of
the suitable photon energies for both single and dual energy γ-ray transmissions
have been investigated. In this work, aluminum concentration values with an
accuracy of 98.53% and 99.09% have been obtained by using both single and
dual γ-ray energy transmission techniques, respectively. A comparison of the
present results with those from the conventional chemical analysis indicates a
fair agreement. A criterion for the best choice of the suitable γ-ray transmission
technique for the aluminum determination has been investigated.

DOI

10.21608/ejs.2000.151484

Volume

23

Article Issue

1

Related Issue

22334

Issue Date

2000-12-01

Receive Date

1999-06-24

Publish Date

2000-12-01

Page Start

147

Page End

158

Print ISSN

1012-5566

Online ISSN

2735-5640

Link

https://ejs.journals.ekb.eg/article_151484.html

Detail API

https://ejs.journals.ekb.eg/service?article_code=151484

Order

14

Publication Type

Journal

Publication Title

Egyptian Journal of Solids

Publication Link

https://ejs.journals.ekb.eg/

MainTitle

Determination of Aluminum in Copper-Aluminum Alloy Using Gamma-Ray Spectrometry

Details

Type

Article

Created At

23 Jan 2023