ABSTRACT
Hip replacement implants are currently increasing worldwide. However, the number
of failing implants is also relatively high. In this paper, the design of a prosthesis hip
joint is analyzed using finite element analysis. A FE model was developed in
accordance with the ISO standard and was verified using ANSYS and SIEMENS
NX packages. Furthermore, FE results were verified using published experimental
results. The verified FE model was then used to simulate the joint under dynamic
loading taking into account the effect of level of activity on scaling the loads up. Two
material types: TI-6AL-4V and Cobalt–chromium alloys were used and the results
showed that TI-6AL-4V alloy proved to have more durability under dynamic loading.
In addition, the results showed that the level and type of life activates have an
obvious and a pronounced effect on the service life of the prosthesis joint.

DOI

10.21608/amme.2018.35024

Keywords

hip implant, hip prosthesis, Finite Element

Authors

First Name

H.

Last Name

Kamel

MiddleName

M.

Affiliation

Egyptian Armed Forces.

Email

-

City

-

Orcid

-

Volume

18

Article Issue

18th International Conference on Applied Mechanics and Mechanical Engineering.

Related Issue

5736

Issue Date

2018-04-01

Receive Date

2019-06-16

Publish Date

2018-04-01

Page Start

1

Page End

12

Print ISSN

2636-4352

Online ISSN

2636-4360

Link

https://amme.journals.ekb.eg/article_35024.html

Detail API

https://amme.journals.ekb.eg/service?article_code=35024

Order

78

Type

Original Article

Type Code

831

Publication Type

Journal

Publication Title

The International Conference on Applied Mechanics and Mechanical Engineering

Publication Link

https://amme.journals.ekb.eg/

MainTitle

MODELING AND SIMULATION OF A HIP PROSTHESIS IMPLANTATION

Details

Type

Article

Created At

22 Jan 2023