Abstract

Diagnostic x-radiation is an essential part of present-day medical practice being the largest contributor of irradiation to the general population. Therefore, routine checking of x-ray tubes leakage is very important to ensure that leakage radiation at a one-meter distance from focus is less than 1mR/h. The measurement of background radiation was performed during normal departmental working hours: early in the morning before the machines were switched on, and after the machines were switched on. Exposures were performed using different exposure factors and the fall out radiation in both control and uncontrolled areas were also recorded. The highest equivalent dose rate was recorded in the uncontrolled area of room 2, where the mean dose rate was 26.21 μSv/h (in changing room) and 11.67 μSv/h (behind the door), respectively. It follows the dose rate measured in the uncontrolled area of room 1, where the mean dose rate was 6.33 μSv/h (behind the door), 2.96 μSv/h (in changing room B), and 1.56 μSv/h (in changing room A), respectively. The lowest radiation dose was measured in both controlled and uncontrolled areas of CT-scan, where the mean dose rate was 0.17 μSv/h and 0.16 μSv/h, respectively, simply because the room doors and wall are adequately lead-lined and protected.  All values obtained in this study are within the permissible limit, except that for uncontrolled areas in x-ray rooms 1 and 2, where radiation dose rates were higher than the reference limit for public exposure.

DOI

10.21608/jntas.2020.23942.1018

Keywords

Radiation, Leakage, Effective dose, controlled area, uncontrolled area

Authors

First Name

Joseph

Last Name

Zira

MiddleName

Dlama

Affiliation

Bayero University Kano Nigeria

Email

josephdlama@gmail.com

City

Kano

Orcid

-

First Name

Uba

Last Name

Zikirullahi

MiddleName

Sani

Affiliation

bayeron University Kano

Email

zakirullahisaniuba@gmail.com

City

kano

Orcid

-

First Name

Idris

Last Name

Garba

MiddleName

-

Affiliation

Bayero University Kano

Email

igarba.radg@buk.edu.ng

City

kano

Orcid

-

First Name

Mohammed

Last Name

Sidi

MiddleName

-

Affiliation

Bayero University Kano

Email

muhammadsidi82@gmail.com

City

kano

Orcid

-

First Name

Mohammed

Last Name

Umar

MiddleName

-

Affiliation

Bayero University Kano

Email

umar.sani888@gmail.com

City

Bauchi

Orcid

0000-0002-0213-0920

First Name

Shem

Last Name

Bature

MiddleName

Silas

Affiliation

Bayero University Kano

Email

shembature2702@gmail.com

City

kano

Orcid

-

Volume

8

Article Issue

1

Related Issue

11080

Issue Date

2020-01-01

Receive Date

2020-02-13

Publish Date

2020-10-18

Page Start

135

Page End

143

Print ISSN

2314-8209

Online ISSN

2314-8217

Link

https://jntas.journals.ekb.eg/article_119428.html

Detail API

https://jntas.journals.ekb.eg/service?article_code=119428

Order

12

Type

Original Article

Type Code

632

Publication Type

Journal

Publication Title

Journal of Nuclear Technology in Applied Science

Publication Link

https://jntas.journals.ekb.eg/

MainTitle

Assessment of Radiation Leakage from Diagnostic Rooms of Radiology Department of a Teaching Hospital in Kano, Northwestern Nigeria