Introduction: Since its introduction, surgical navigation systems have become integrated into surgical practices that include complex nasal endoscopic procedures to avoid serious complication such as injury to the optic nerve, the carotid artery, the dura mater and the brain parenchyma.
Objective: We aimed to determine the accuracy of image guided system (IGS) by using different timing interval of computed tomography scan of paranasal sinus and our experience in performing the surface marking.
Methods: A cross sectional study was done. Subjects were recruited from patients planned for endoscopic sinus surgery with IGS, who met the inclusion and exclusion criteria. The image guided navigation measurements were compared to the actual anatomical measurements of intercanthal distance of both eyes and the tangential line of inverted triangle to the philtrum. The estimated time to register the IGS was recorded. All measurements were done at the same time as the calibration of the IGS.
Results: The mean intercanthal distance measured by IGS was 38.17 mm (SD 3.34) whereas the actual intercanthal distance was 38.17 mm (SD 3.37 mm). The tangential line measured by IGS was 46.62 (SD 3.39) whereas the actual measurement of the tangential line was 46.47mm (SD 3.27). The was no significant difference in both measurements between IGS and actual anatomical measurements (p‌=0.804 and 0.496, respectively).
Conclusions: Even with the different intervals there was not much deviation from the actual anatomical distance. The IGS is accurate and is applicable for use in Asian patients.

Keywords

Electromagnetic navigation system, Endoscopic sinus surgery, registration accuracy, surface marking

Authors

First Name

Zabrina

Last Name

Samarakkody

MiddleName

Marnel

Affiliation

Department of Otorhinolaryngology-Head & Neck Surgery, School of Medical Sciences, Universiti Sains Malaysia

Email

zabrinagill@gmail.com

City

-

Orcid

-

First Name

Harvinder

Last Name

Sran

MiddleName

Singh

Affiliation

Department of Otorhinolaryngology, Hospital Raja Permaisuri Bainun, Malaysia

Email

-

City

-

Orcid

-

First Name

Norasnieda

Last Name

Shukri

MiddleName

Md

Affiliation

Department of Otorhinolaryngology-Head & Neck Surgery, School of Medical Sciences, Universiti Sains Malaysia,

Email

-

City

-

Orcid

-

First Name

Baharudin

Last Name

Abdullah,

MiddleName

-

Affiliation

Department of Otorhinolaryngology Head and Neck Surgery, School of Medical Sciences, Universiti Sains Malaysia, Kelantan

Email

profbaha@gmail.com

City

-

Orcid

-

Volume

20

Article Issue

2

Related Issue

6847

Issue Date

2019-07-01

Receive Date

2019-01-11

Publish Date

2019-07-01

Page Start

67

Page End

73

Print ISSN

2090-0740

Online ISSN

2090-3405

Link

https://ejentas.journals.ekb.eg/article_43946.html

Detail API

https://ejentas.journals.ekb.eg/service?article_code=43946

Order

4

Type

Original Article

Type Code

467

Publication Type

Journal

Publication Title

Egyptian Journal of Ear, Nose, Throat and Allied Sciences

Publication Link

https://ejentas.journals.ekb.eg/

MainTitle

Registration accuracy and surface marking of an electromagnetic navigation system for use in endoscopic sinus surgery

Details

Type

Article

Created At

22 Jan 2023