Abstract
Purpose
Function formulation to estimate the most appropriate threshold for automatic lesion delineation in PET scan.
Materials and Methods
The phantom consists of two parts: A hollow cylinder with set of spheres. The sphere assembly contains 4 spheres of different volumes. The phantom was filled with water containing 18F-Fluorodeoxyglucose (18F-FDG) forming the sphere versus background activity concentration (μCi/ml) ratios 3:1, 5:1, 8:1, 10:1, 12:1, 14:1 and15:1. Graphs were plotted, of standardized uptake value threshold versus lesion-size at different sphere/background ratio. There was a logarithmic decrease in threshold with an increase in lesion size. The most appropriate SUV threshold for each lesion at different contrast ratio was estimated from the plotted graphs. Four tests were used to generate a 3D general function to estimate threshold using the contrast ratio and lesion volume as inputs and the other three tests were used for the function validation. The Normalized mean square error (NMSE) is used to measure the convergence between estimated and measured data set.
Results
The created logarithmic functions offer an easy and simple way to estimate the percentage value of SUVmax as a threshold using the lesion/background ratio and sphere volume(or Diameter) from CT images as inputs to be used for lesion delineation and volume determination in PET images that was verified and the NMSE=6.2
Conclusion:
The created formula is a useful and easy to estimate the most appropriate SUV threshold for automatically definite lesion volume.

DOI

10.21608/ajnsa.2019.3449.1080

Keywords

positron emission tomography-computed tomography, standardized uptake value, lesion delineation

Authors

First Name

Hesham

Last Name

Ahmed

MiddleName

Abdel Gawad

Affiliation

Kasr Al Aini center for nuclear medicine and oncology

Email

hesham_ag@yahoo.com

City

Cairo

Orcid

-

First Name

Shaban

Last Name

Al Ramlawy

MiddleName

-

Affiliation

Faculty of Medicine Department of Critical Care- Cairo University Hospitals Cairo University, Egypt

Email

shabanwmr@hotmail.com

City

-

Orcid

-

First Name

Magdy

Last Name

Khalil

MiddleName

-

Affiliation

Helwan University Department of Physics (Medical Biophysics), Faculty of Science Cairo Egypt

Email

magdy_khalil@hotmail.com

City

-

Orcid

-

Volume

52

Article Issue

2

Related Issue

4364

Issue Date

2019-04-01

Receive Date

2018-05-07

Publish Date

2019-04-01

Page Start

169

Page End

180

Print ISSN

1110-0451

Online ISSN

2090-4258

Link

https://ajnsa.journals.ekb.eg/article_29303.html

Detail API

https://ajnsa.journals.ekb.eg/service?article_code=29303

Order

21

Type

Original Article

Type Code

455

Publication Type

Journal

Publication Title

Arab Journal of Nuclear Sciences and Applications

Publication Link

https://ajnsa.journals.ekb.eg/

MainTitle

Estimation of the most Appropriate Threshold for Lesion Delineation and Volumetric Quantification in 18F-FDG PET

Details

Type

Article

Created At

22 Jan 2023