5493

Contrast Evaluation for a Dual Head SPECT System with Different Energy PeakingDrift

Article

Last updated: 03 Jan 2025

Subjects

-

Tags

-

Abstract

Gamma cameras contain energy discriminators that allow only those photons within a specified energy range to be recorded. The process of setting this energy window is called “peaking" the camera. A spontaneous shift in the peak of one head of a triple-head gamma camera was reported to cause an artifact that mimicked ischemia on 201Tl-chloride myocardial SPECT. The characteristic energy distribution results from the collection of total photon energy absorbed by the detector. Physically peaked detectors will center the energy window on the photo peak of selected isotope. Aim of the study: Our aim is to evaluate the effect of energy peak drifting for both imaging heads on the image contrast in SPECT. Materials and methods: Using a dual head gamma camera, the SPECT phantom with its inserts was scanned using a fixed acquisition protocol and image reconstruction parameters but with changing the energy peak window adjustment which was operated for both heads at different values starting from 120 KeV up to 160 KeV with incremental step of 5 KeV, and the window width was adjusted to be 15% as used in the clinical applications.
Results: Our results shows a fluctuation in the image contrast, as there was a gradual increase in the contrast value from 120 KeV setting up to 140 KeV by 31 % and also there was a gradual decrease in the contrast value from 140 KeV setting up to 160 KeV setting by 15 %. The clinical verification for these results was done on cardiac patient who underwent a cardiac scan twice with the normal gamma camera peak setting and then with off-peak setting at 120 KeV (which gives the lowest contrast value for the phantom study). And the case interpretation results confirm the phantom results as there was a significant change in the uptake in the left ventricle wall, with also a verification from the quantitative results by the Quantitave comparison using the quantitative gated SPECT data, that shows a 9% difference in the EF calculated value from the normal to the off-peak setting, which may affect the interpretation and in consequence the further treatment of the patient.
Conclusion: The energy window peaking for the gamma camera has a proven effect on the planar images, and this was explained and proved in our study to have a significant effect on the contrast of the SPECT images which will affect the interpretation of the clinical results.

DOI

10.21608/egyjnm.2011.5493

Authors

First Name

Ibrahim

Last Name

Saad Ahmed

MiddleName

Elsayed

Affiliation

Department of Clinical Oncology and Nuclear medicine- Faculty of Medicine- Cairo University.

Email

i_elsayed@cu.edu.eg

City

-

Orcid

0000-0001-5829-5196

Volume

4

Article Issue

4

Related Issue

579

Issue Date

2011-12-01

Receive Date

2018-02-21

Publish Date

2011-12-01

Page Start

58

Page End

67

Print ISSN

1687-4994

Online ISSN

2536-9113

Link

https://egyjnm.journals.ekb.eg/article_5493.html

Detail API

https://egyjnm.journals.ekb.eg/service?article_code=5493

Order

8

Type

Original Paper, Physics

Type Code

387

Publication Type

Journal

Publication Title

Egyptian Journal Nuclear Medicine

Publication Link

https://egyjnm.journals.ekb.eg/

MainTitle

Contrast Evaluation for a Dual Head SPECT System with Different Energy PeakingDrift

Details

Type

Article

Created At