A Comparison between Two Methods of Performing the Daily Intrinsic Uniformity Quality Control Test for Gamma Camera

Article

Last updated: 22 Jan 2023

Subjects

-

Tags

-

Abstract

Field flood uniformity is a fundamental prerequisite for using the gamma camera in clinical routine. Different techniques are available for evaluating the intrinsic flood uniformity of dual head gamma camera during the daily quality control. Some manufacturers recommend a simultaneous acquisition (manufacturer method) using the two detectors and others use a single detector method (conventional method). A comparison of the two methods is questioned to determine how much difference might be seen if one of the two methods is applied routinely. Two methods were used to evaluate the intrinsic uniformity of a commercially available gamma camera (Siemens ECAM, Dual head) applying the following acquisition parameters: number of acquired counts, count rates, source volume, offset center x-axis, offset center y-axis and source-tocamera distance. One method uses the two detectors simultaneously (denoted by A) while the other method uses a one detector at a time (denoted by B). The integral and differential uniformity for the central and useful field of view were calculated and recorded for comparison. The acquisition parameters that achieved the optimal uniformity values are tabulated below. Method A demonstrated better uniformity values over that yielded by the conventional method B with respect to all acquisition parameters considered (p<0.05). Using the tabulated results, a daily routine of flood uniformity will take approximately a time of 7-10 min if method A is used while a time of 10-12 min is taken if the conventional single detector method is used.

DOI

10.21608/egyjnm.2010.5499

Authors

First Name

Gaber

Last Name

Ziada

MiddleName

-

Affiliation

Professor of Nuclear Medicine physics, Kuwait University.

Email

gz555@hotmail.com

City

-

Orcid

-

Volume

3

Article Issue

3

Related Issue

577

Issue Date

2010-12-01

Receive Date

2018-02-21

Publish Date

2010-12-01

Page Start

40

Page End

54

Print ISSN

1687-4994

Online ISSN

2536-9113

Link

https://egyjnm.journals.ekb.eg/article_5499.html

Detail API

https://egyjnm.journals.ekb.eg/service?article_code=5499

Order

5

Type

Original Paper, Physics

Type Code

387

Publication Type

Journal

Publication Title

Egyptian Journal Nuclear Medicine

Publication Link

https://egyjnm.journals.ekb.eg/

MainTitle

-

Details

Type

Article

Created At

22 Jan 2023