Abstract

A simple and efficient established method that is characterized by its quickness and high sensitivity for determining ferrous ion in pure and pharmaceutical formulations by the generate of dark-blue precipitate using reaction of hexacyanoferrate(III) with ferrous ion in ammonium acetate medium. The equipment NAG-4SX3-3D analyzer was used to quantify turbidity by measuring the incident light attenuation crashing on the precipitate surface granules (0-180 degree). The optimum parameter has been investigated in order to improve the newly created approach's sensitivity.

The linear range (0.1-7) millimol.L-1 with (r=0.9971), (correlation coefficient), percentage linearity (R % = 99.42), and RSD % for the repetition (n=7) also were lower than 0.1 percent for the measurement of ferrous ion, with concentration (1, 4.7 millimol .L-1). Whereas L.O.D. = 212.6712 ng/sample from the progressive dilution throughout the calibration graph's lowest concentration linear dynamic range the suggested approach was compared to the traditional method (UV- spectrophotometric at λmax = 222 nm of analysis employing the conventional technique of adding by the use of paired t- test (the mean difference between pairs of measurements). It's notice that with calculated numbers, there wasn't any substantial variation between the values given of any particular firm. t-value at 95% interval from the proposed approach.

DOI

10.21608/ejchem.2022.131307.5788

Keywords

Fe (II), Turbidity, Flow Injection Analysis

Authors

First Name

Sarah

Last Name

Hameed

MiddleName

-

Affiliation

Department of chemistry, College of science, University of Baghdad, Baghdad, Iraq

Email

sarahf.hameed@yahoo.com

City

-

Orcid

-

First Name

Nagham

Last Name

Turkey

MiddleName

S

Affiliation

Chemistry, Faculty of Science, University of Baghdad, Baghdad, Iraq

Email

nagamturkey@yahoo.com

City

-

Orcid

-

Volume

65

Article Issue

13

Related Issue

37459

Issue Date

2022-12-01

Receive Date

2022-04-03

Publish Date

2022-12-01

Page Start

871

Page End

886

Print ISSN

0449-2285

Online ISSN

2357-0245

Link

https://ejchem.journals.ekb.eg/article_237834.html

Detail API

https://ejchem.journals.ekb.eg/service?article_code=237834

Order

237,834

Type

Original Article

Type Code

297

Publication Type

Journal

Publication Title

Egyptian Journal of Chemistry

Publication Link

https://ejchem.journals.ekb.eg/

MainTitle

Determination of Ferrous Ion in Pure & Pharmaceutical Preparation by Continuous Flow Injection Analysis Via Turbidmetric Utilizing NAG-4SX3-3D Analyzer

Details

Type

Article

Created At

22 Jan 2023