Beta
69942

Measurements of Qf СйП, fmsr by Using Different Detectors in Radiation Qclin, Qmse Therapy Systems

Article

Last updated: 01 Jan 2025

Subjects

-

Tags

-

Abstract

Abstract
Background: It worth to be mentioned that the use of small size photon beam is frequently used in modern radio-therapy to treat brain tumors and functional disorders.
Aim of Study: To measure output factors (OFs) and cal-culate the field factor ( ƒclin, ƒmsr ) of small fields and
V clin, Qmse
study its suitability for small field relative dosimetry.
Material and Methods: Numerous detectors were used for measuring the output factors for 6 MV photon beams by a CyberKnife®. To normalize different detector responses for the same field configuration, a correction factor was calculated for each detector by simulating the radiation delivery using Monte Carlo (MC) methods. Detectors used in the study were PTW60019 MicroDiaond, PTW 60018 Silicon diode, PTW31018 MicroLion, and Exradin W1 Scintillator. Field factors were calculated using Alfonso formula. Output factors for a CyberKnife were measured in circular fields with the diameters range from 5mm to 60mm. Measurements were made in a water tank at a 1.5cm in term of depth and at 80 cm for source-to-axis distance.
Results: The results of the current study show that the output measured by the detectors Micro Diamond and Exradin W1 Scintillator (PSD) were within the uncertainties of the Monte Carlo simulations for all the beam cones. The silicon diode detector was over-responding, while the MicroLion was under-responding. We found that an accurate dosimeter could be the MicroDiamond and Exradin W1 Scintillators in small field dosimetry.
Conclusion: We found that at small collimator settings only the synthetic microdiamond (PTW60019) and Exradin W1 scintillation detectors can be used as relative dosimeters without applying any correction factor.

DOI

10.21608/mjcu.2019.69942

Keywords

Synthetic diamond, Exradin W1 Scintillator (PSD), Small field dosimetry

Authors

First Name

SAMAH K. ELSHAMNDY, M.Sc.;

Last Name

PAOLO FRANCESCON, Ph.D.

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

NINFA SATARIANO, Ph.D.;

Last Name

CHIARA ORLANDI, Ph.D.

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

First Name

HASSAN ABOUELENEIN, Ph.D.;

Last Name

AHMED M. AHMED, Ph.D.

MiddleName

-

Affiliation

The Department of Radiotherapy and Nuclear Medicine, Faculty of Medicine, Sohag University, Sohag, Egypt

Email

-

City

-

Orcid

-

Volume

87

Article Issue

September

Related Issue

8905

Issue Date

2019-09-01

Receive Date

2019-02-04

Publish Date

2019-09-01

Page Start

3,739

Page End

3,747

Print ISSN

0045-3803

Online ISSN

2536-9806

Link

https://mjcu.journals.ekb.eg/article_69942.html

Detail API

https://mjcu.journals.ekb.eg/service?article_code=69942

Order

123

Type

Original Article

Type Code

263

Publication Type

Journal

Publication Title

The Medical Journal of Cairo University

Publication Link

https://mjcu.journals.ekb.eg/

MainTitle

Measurements of Qf СйП, fmsr by Using Different Detectors in Radiation Qclin, Qmse Therapy Systems

Details

Type

Article

Created At

22 Jan 2023