Aim of the study: The current study was conducted to determine the accuracy of QuickScan imaging protocols of iCAT Flx CBCT machine in detection and measuring crestal bone defects around dental implants.
Materials and Methods: 20 simulated peri-implant crestal bone defects were created in bovine ribs bone models then were scanned by iCAT FLX CBCT machine with HD Scan 0.25 mm voxel, QuickScan 0.25mm voxel, and QuickScan + 0.3mm voxel sizes. All scans were imported in on-demand 3D App software and produced linear measurements were compared to true measurements done directly with digital caliper on bone models.
Results: All imaging protocols of CBCT showed no statistical significance in detection of shallow peri-implant bone defects while the QuickScan + imaging protocol showed less accuracy in measuring the depth of defects.
Conclusion: QuickScan + can be used as a low dose periodic follow up imaging utility for postoperative implant surgery for detection of any small bone changes but if more details are needed QuickScan technique is required.

DOI

10.21608/edj.2018.76810

Keywords

Peri-implant defect, QuickScan, CBCT, linear accuracy

Authors

First Name

Wael

Last Name

Aboelmaaty

MiddleName

-

Affiliation

Lecturer of Oral Radiology and Diagnosis, Faculty of Dentistry, Mansoura University

Email

-

City

-

Orcid

-

First Name

Nermin

Last Name

Morgan

MiddleName

-

Affiliation

Assistant Lecturer of Oral Radiology and Diagnosis, Faculty of Dentistry, Mansoura University

Email

-

City

-

Orcid

-

First Name

Ehab

Last Name

Abdelfadil

MiddleName

-

Affiliation

Lecturer of Oral and Maxillofacial Surgery, Faculty of Dentistry, Mansoura University

Email

-

City

-

Orcid

-

First Name

Mostafa

Last Name

Ashmawy

MiddleName

-

Affiliation

Lecturer of Oral Radiology, Oral Radiology Department, Faculty of Dentistry, Ain-Shams University

Email

-

City

-

Orcid

-

Volume

64

Article Issue

Issue 3 - July (Oral Medicine, X-Ray, Oral Biology & Oral Pathology)

Related Issue

11453

Issue Date

2018-07-01

Receive Date

2020-03-10

Publish Date

2018-07-01

Page Start

2,327

Page End

2,336

Print ISSN

0070-9484

Online ISSN

2090-2360

Link

https://edj.journals.ekb.eg/article_76810.html

Detail API

https://edj.journals.ekb.eg/service?article_code=76810

Order

23

Type

Original Article

Type Code

254

Publication Type

Journal

Publication Title

Egyptian Dental Journal

Publication Link

https://edj.journals.ekb.eg/

MainTitle

Accuracy of QuickScan Imaging Protocols of iCAT FLX CBCT in Assessment of Peri-implant Bone Defects

Details

Type

Article

Created At

22 Jan 2023