Beta
329747

Accuracy and precision of non-invasive continuous haemoglobin concentration monitoring in diabetic patients

Article

Last updated: 05 Jan 2025

Subjects

-

Tags

-

Abstract

Background
Accuracy and precision of non-invasive continuous haemoglobin concentration (SpHb) provided by Masimo device in diabetic patients is poorly studied. This retrospective analysis aimed to provide data on SpHb accuracy and precision in diabetic patients.
Results
The sample size population consisted of 14 patients, with 56 SpHb/Lab data pairs. Lab value showed a mean ± standard deviation (SD) of 13.2 ± 1.2 g/dL, whilst SpHb showed a mean ± SD of 11.8 ± 1.1 g/dL. Linear regression analysis between Lab/SpHb data pairs showed a of 0.8960 (CI 0.8281-0.9379, value < 0.0001). SpHb underestimated the real Hb values provided by Lab. Bland-Altman analysis showed that SpHb accuracy was −1.37 g/dL (CI −1.51 to −1.22 g/dL, value < 0.0001), precision of 0.55 g/dL, lower LOA −2.45 g/dL (CI −2.71 to −2.20 g/dL) and upper LOA −0.28 g/dL (CI −0.53 to −0.02 g/dL).
Conclusions
For the first time, we provided data on SpHb accuracy and precision in the diabetic population. SpHb showed a high correlation coefficient when compared with Lab values, but the wide LOA limits its accuracy.

DOI

10.1186/s42077-021-00201-2

Keywords

haemoglobin, Non-invasive monitoring, diabetes, Aortic surgery

Authors

First Name

Rosanna Carmela

Last Name

De Rosa

MiddleName

-

Affiliation

-

Email

rosderosa.rdr@gmail.com

City

-

Orcid

-

First Name

Antonio

Last Name

Romanelli

MiddleName

-

Affiliation

-

Email

-

City

-

Orcid

-

Volume

14

Article Issue

1

Related Issue

44741

Issue Date

2022-01-01

Receive Date

2021-12-09

Publish Date

2022-01-10

Print ISSN

1687-7934

Online ISSN

2090-925X

Link

https://asja.journals.ekb.eg/article_329747.html

Detail API

https://asja.journals.ekb.eg/service?article_code=329747

Order

329,747

Publication Type

Journal

Publication Title

Ain-Shams Journal of Anesthesiology

Publication Link

https://asja.journals.ekb.eg/

MainTitle

Accuracy and precision of non-invasive continuous haemoglobin concentration monitoring in diabetic patients

Details

Type

Article

Created At

20 Dec 2024