Beta
218140

Implant Surgical Guide Accuracy: CBCT cast scan

Article

Last updated: 28 Dec 2024

Subjects

-

Tags

Healthcare research

Abstract

The aim of this study is to evaluate the accuracy of implant placement using implant 
surgical guide by CBCT cast scan. Patients and methods: seven resin surgical guide models were 
used in this study. Implant position was virtually planned using 3D planning software. Surgical guides 
were designed on the software and exported in STL format. Ten implants were placed. The 3D 
position of the planned and placed implants, in terms of the linear deviations of the implant head and 
apex and the angular deviations of the implant axis, was compared by superimposing the pre- and 
postoperative CBCT using software. Data showed parametric distribution so; it was represented by 
mean and standard deviation (SD) values. Result: Angular deviation was 2.38 ± 0.37 degree. Coronal 
deviation of Implant was 0.49 ± 0.12 mm. Apical deviation of implant Apex was 1.28 ± 0.15 mm, 
although the implant position was slightly different than the planning but, there was no statistically 
significant difference. Conclusion Implant surgical guide fabricated with CBCT cast scan has 
accepted accuracy and it is less costly technique with accepted accuracy. 

DOI

10.21608/mjmr.2022.218140

Keywords

accuracy, CBCT, dental cast model, Surgical Guide

Authors

First Name

Mahmoud .

Last Name

Elmadany

MiddleName

A

Affiliation

Department of Oral and maxillofacial Radiology, Faculty of Dentistry, South Valley University, Egypt.

Email

-

City

-

Orcid

-

First Name

Maha .

Last Name

Amer

MiddleName

E

Affiliation

Department of Head of Oral and maxillofacial Radiology, Faculty of Dentistry, Minia University, Egypt.

Email

-

City

-

Orcid

-

First Name

Emad

Last Name

Tolba

MiddleName

M.

Affiliation

Department of Head of Oral and maxillofacial Radiology, Faculty of Dentistry, Minia University, Egypt.

Email

-

City

-

Orcid

-

First Name

Mahmoud Mahmoud Agamy

Last Name

Mahmoud Agamy

MiddleName

-

Affiliation

Department of Prosthodontics, Faculty of Dentistry, Minia University, Egypt

Email

-

City

-

Orcid

-

Volume

31

Article Issue

4

Related Issue

31304

Issue Date

2020-10-01

Receive Date

2022-02-08

Publish Date

2020-10-01

Page Start

282

Page End

286

Online ISSN

2682-4558

Link

https://mjmr.journals.ekb.eg/article_218140.html

Detail API

https://mjmr.journals.ekb.eg/service?article_code=218140

Order

218,140

Type

Original Article

Type Code

2,212

Publication Type

Journal

Publication Title

Minia Journal of Medical Research

Publication Link

https://mjmr.journals.ekb.eg/

MainTitle

Implant Surgical Guide Accuracy: CBCT cast scan

Details

Type

Article

Created At

23 Jan 2023