Beta
146290

COMPARING HEARING AID FITTING OF THE MANUFACTURER'S FIRST FIT VERSUS NAL-NL2 USING REAL EAR MEASUREMENTS

Article

Last updated: 04 Jan 2025

Subjects

-

Tags

-

Abstract

Real ear measurments (REMs) or the probe microphone measurments is an objective method to verify hearing aids fitting taking into account the ear canal acoustics of the patient.(1) It is the best method to provide a patient with an audible signal.(2)
Prescriptive fitting formulas, such as NAL-NL1, NAL-NL2, and DSL V5a are evidence-based methods for determining how much amplification is needed for a particular hearing loss. By using prescriptive formula verified with REMs at the initial hearing aid fitting allows the audiologist to ensure audibility for conversational speech, to set appropriate amounts of amplification for soft sounds, and to prevent loud sounds from becoming uncomfortable.(3)
Some audiologists use the first-fit formulas of the manufacturer to have a proper fitting gain, however, many studies revealed that the first fit of the manufacturer is insufficient for the required gain.(4)
Self-report outcome measure is also mandatory to determine the benefit of hearing aids.(5) The Abbreviated Profile of Hearing Aid Benefit questionnaire (APHAB) is used to measure disability caused by hearing loss and measure the reduction of that disability by using hearing aids. The APHAB uses 24 elements including 4 dimensions: ease of communication, reverberation, background noise, and aversiveness to sounds.(6)

DOI

10.21608/alexpo.2021.60879.1116

Keywords

Real ear measurments, MANUFACTURER'S FIRST FIT, APHAB

Authors

First Name

Heba

Last Name

Sharaky

MiddleName

Gamal Eldin

Affiliation

aAudio-vestibular Medicine Unit, Department of Otorhinolaryngology, Faculty of Medicine, Alexandria University, Alexandria, Egypt.

Email

h_gamaleldin16@alexmed.edu.eg

City

-

Orcid

-

Volume

3

Article Issue

1

Related Issue

22925

Issue Date

2021-03-01

Receive Date

2021-02-01

Publish Date

2021-03-01

Page Start

21

Page End

22

Online ISSN

2682-2636

Link

https://alexpos.journals.ekb.eg/article_146290.html

Detail API

https://alexpos.journals.ekb.eg/service?article_code=146290

Order

21

Type

Preliminary preprint short reports of original research

Type Code

1,426

Publication Type

Journal

Publication Title

ALEXMED ePosters

Publication Link

https://alexpos.journals.ekb.eg/

MainTitle

COMPARING HEARING AID FITTING OF THE MANUFACTURER'S FIRST FIT VERSUS NAL-NL2 USING REAL EAR MEASUREMENTS

Details

Type

Article

Created At

23 Jan 2023